The IKA® calorimeter C 5000 control is the only calorimeter in the world that offers a free selection of 3 working methods:
Thus it is possible to perform determinations of gross calorific values of liquid and solid samples in adiabatic (approx. 14 – 18 min), isoperibolic (approx. 22 min) and dynamic (reduced time: approx. 10 min) mode. A high level of automation in addition to an extensive range of accessories leave nothing more to wish for.
– Automatic water handling system includes tempering, filling and emptying of calorimeter inner vessel
– Automatic oxygen filling and degassing of the decomposition vessel
– Automatic decomposition vessel identification
– Validation according to DIN 51900,ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711
– Interface connections for each of the following: scale, printer, monitor and sample rack C 5020
– User-friendly software C 5040 CalWin for controlling the calorimeter and administration of measuring data (accessory)
– LIMS integration is possible
– Cooling water supply via thermostat f.e IKA KV 600 ( accessory ) or firmly installed water connection, (temperature range 10 – 19°C )
– Special halogen resistent vessel C 5012 for quantitative decomposition of halogens and sulfur (accessory)
– The decomposition vessel can be changed over to use combustible crucibles C 14 (accessory C 5010.4 is needed)
– Standard consumables for calibrations and initial operation included with delivery
Technical Data:
Measuring range max. | 40000 J |
Measuring mode adiabatic 22°C | On Request |
Measuring mode dynamic 22°C | On Request |
Measuring mode isoperibol 22°C | On Request |
Measuring mode adiabatic 25°C | yes |
Measuring mode dynamic 25°C | yes |
Measuring mode isoperibol 25°C | yes |
Measuring mode dynamic 30°C | no |
Measuring mode isoperibol 30°C | no |
Measuring mode double dry (ISO 1928) | no |
Measuring time adiabatic approx. | 15 min |
Measuring time dynamic approx. | 10 min |
Measuring time isoperibol approx. | 22 min |
Reproducibility adiabatic (1g benzoic acid NBS39i) | 0.05 %RSD |
Reproducibility dynamic (1g benzoic acid NBS39i) | 0.1 %RSD |
Reproducibility isoperibol (1g benzoic acid NBS39i) | 0.05 %RSD |
Working temperature max. | 25 °C |
Temperature measurement resolution | 0.0001 K |
Cooling medium temperature min. | 10 °C |
Cooling medium temperature max. | 20 °C |
Cooling medium permissible operating pressure | 9 bar |
Cooling medium | tap water |
Type of cooling | flow |
Flow rate min. | 15 l/h |
Flow rate max. | 60 l/h |
Rec. flow rate at 18°C | 40 l/h |
Oxygen operating pressure max. | 40 bar |
Interface scale | RS232 |
Interface printer | Centronix |
Interface PC | RS232 |
Interface test rack | yes |
Interface ext. monitor | no |
Interface ext. keyboard | no |
Oxygen filling | yes |
Degasification | yes |
Decomposition detection | yes |
Decomposition vessel C 5010 | yes |
Analysis according to DIN 51900 | yes |
Analysis according to ASTM D240 | yes |
Analysis according to ASTM D4809 | yes |
Analysis according to ASTM D1989 | yes |
Analysis according to ASTM D5468 | yes |
Analysis according to ASTM D5865 | yes |
Analysis according to ASTM E711 | yes |
Dimensions (W x H x D) | 560 x 400 x 380 mm |
Weight | 40 kg |
Permissible ambient temperature | 20 – 25 °C |
Permissible relative humidity | 80 % |
Protection class according to DIN EN 60529 | IP 21 |
RS 232 interface | yes |
Voltage | 230 V |
Frequency | 50/60 Hz |
Power input | 1060 W |