HI96800
$100.00
HANNA – HI96800
Digital Refractometer for Refractive Index and Brix
Hanna’s new HI96800 Digital Refractometer measures Refractive Index and displays results in both temperature compensated (nD20) and non-temperature compensated (nD) formats on the dual-level LCD screen. The HI96800’s intuitive layout and pinpoint accuracy ensures effortless operation and reliable readings every time. Refractive Index is measured to determine a liquid’s composition and assess its concentration and density. It is used by many industries including pharmaceutical, food and environmental, as an important indicator of quality and consistency. The HI96800 also converts Refractive Index to % Brix with the push of a single button.
Sample size as small as 2 metric drops (100 μl)
Sealed stainless steel well with high-grade optical prism made of flint glass
Fast 1.5 second response time for temperature compensated readings
Add to cart
Product Description
HANNA – HI96800
Digital Refractometer for Refractive Index and Brix
Specifications:-
Sugar Content Range | 0.0 to 85.0% Brix; 1.3300 to 1.5080 nD; 1.3330 to 1.5040 nD₂₀ |
Sugar Content Resolution | 0.1 % Brix; 0.0001 nD; 0.0001 nD₂₀ |
Sugar Content Accuracy (@25ºC/77ºF) | ±0.2% Brix; ±0.0005 nD; ±0.0005 nD₂₀ |
Temperature Range | 0 to 80°C (32 to 176°F) |
Temperature Resolution | 0.1°C (0.1°F) |
Temperature Accuracy | ±0.3 °C (±0.5 °F) |
Temperature Compensation | automatic between 10 and 40°C (50 to 104°F) |
Measurement Time | approximately 1.5 seconds |
Minimum Sample Volume | 100 μL (to cover prism totally) |
Light Source | yellow LED |
Sample Cell | stainless steel ring and flint glass prism |
Auto-off | after three minutes of inactivity |
Enclosure Rating | IP65 |
Battery Type / Life | 9V / approximately 5000 readings |
Dimensions / Weight | 192 x 102 x 67 mm (7.6 x 4.01 x 2.6”) / 420 g (14.8 oz.) |
Ordering Information | HI96800 is supplied with battery and instruction manual |