Ultimate 3000 UHPLC System

Ultimate 3000 UHPLC System

$100.00

Thermofisher-Ultimate-3000-UHPLC-system

The UltiMate WPS-3000TXRS Rapid Separation Well Plate Autosampler with Extended Pressure Range offers:

  • Support of up to 12 well plates
  • Injection cycle times as low as 20 seconds for fast applications
Ultimate 3000UHPLC

$100.00

Add to cart
Category:

Product Description

Thermofisher-Ultimate-3000-UHPLC-System

Open architecture for easy sample access and customization
Both Open Autosampler configurations provide outstanding sample capacity for high-throughput applications.

The WPS-3000TXRS accommodates a maximum of six normal or deep well plates, 6´54 2-mL vials or any combination of these formats.

The WPS-3000TXRS can handle up to 12 normal well plates.

Description
Quaternary Analytical Pump LPG-3400SD
Operating principle :
Serial dual-piston
Gradient Formation: 
Low-pressure gradient proportioning
Recommended flow rate range : (settable range)
0.2–10 mL/min (0.001–10 mL/min)
Flow rate accuracy:
±0.1%
Flow precision :
< 0.05% RSD or < 0.01 min SD, whichever is greater
Pressure range :
2–62 MPa (290–9,000 psi)
Pulsation Typically:
< 0.2 MPa or < 1% whichever is greater
Proportioning accuracy:
±0.5% (of full scale)
Proportioning precision:
< 0.15% SD
No. of eluent lines : 4
Gradient delay volume :690 μL
Solvent degassing : Built-in, 4-channels
 Power Cord India Version,1.8m
Solvent Rack SR-3000
ACC-3000T Autosampler Column Compartment with sample thermostatting
Sample capacity:
3 × 40 vials (1.8 mL) or
3 × 72 vials (1.2 mL) or
3 × 22 vials (4 mL) or
3 × 10 vials (10 mL)
plus 3 × 5 vials (10 mL)
Injection technique:
Pulled loop, with programmable needle wash
Injection methods :
Pulled loop, using full loop or partial loop fill
Injection volume precision:
<0.25% RSD in full loop mode (20 µL loop)
Typically :<0.3% RSD in partial loop mode at 5 µL (20 µL loop)
Carry over :<0.02% with caffeine
Injection cycle time :Typically <45 seconds with 5 µL injection
Sample thermostatting :8–40 °C, min. 15 °C below ambient
Column temperature range: 5 °C above ambient to 50 °C
Maximum operating pressure : 62 MPa (9000 psi)
Sample loop 200uL
1000 ul Syringe
Diode Array Detector DAD-3000
Optical System:
Single-beam, reverse optics design with concave holographic grating
Maximum Data Collection Rate : 100 Hz: DAD-3000
Wavelength Range :190 to 800 nm
No. of Diodes : 1024
Noise Wide slit: < ±8µAU at 254 nm
Narrow slit:
< ±10 µAU at 254 nmresponse time 2 s (according to ASTM time constant approx 1 s), 4 nm bandwidth, water at 1.0 mL/min
Drift :
<1 mAU/h (typically <0.5 mAU/h) at 254 nm and 520 nm with water at 1.0 mL/min
Linearity:
<3% RSD and corr. coeff. >0.9995 up to 1.5 AU,
typically <5% RSD and corr. coeff. ≥0.999 up to 2.0 AU
Light Source :
Deuterium lamp, tungsten lamp, temperature control for both lamps
Wavelength Accuracy :
±1.0 nm, self-calibration with D-alpha line, verification with holmium oxide filter
Pixel resolution :<1 nm
3 D Option        (required for DAD Detector )
Chromeleon CM-CHM-1.
Chromeleon Chromatography data system for control, acquisition, processing, & Reporting software.
32-bit software design for Windows XP Professional/2000
Flexible Reporting as users desires for complete
chromatography information Reporting.